联系电话: 027-87056722
锁相放大器
SR865数字锁相放大器
产品概述和特点 技术参数 相关下载

mHz to 2 MHz (operates to 2.5 MHz)

Low noise voltage and current inputs

1 µs to 30 ks time constants

High bandwidth outputs

Touchscreen data display - large numeric results, chart recordings, & FFT displays

10 MHz timebase input and output

GPIB, RS-232, Ethernet and USB

HDMI video output

 

Signal Channel

Voltage inputs

Single-ended (A) or differential (A-B)

Sensitivity (output scale)

nV to 1 V (voltage input)
fA to 1 µA (current input)

Voltage Input range

10 mV to 1 V (peak).

Current input range

1 µA or 10 nA (peak)

Max input before overload

1 V (peak) or 1 µA (peak)

Input impedance

 

     Voltage input

10 MΩ + 25 pF, AC (>1 Hz) or DC coupled

     Current input

 or 100 Ω to virtual ground

Gain accuracy

±1 % (<200 kHz), ±2 % (to 2 MHz). Signal amplitude must be less than 30% of input range.

Noise

2.5 nV/√Hz at 1 kHz (10 mV input range, typ.)

Harmonic distortion

-80 dB (<100 kHz), -60 dB (>100 kHz)

CMRR

90 dB at 1 kHz (DC coupled)

Dynamic reserve

120 dB (typ)

Reference Channel

Frequency range

0.001 Hz to 2 MHz, specified. (operates to 2.5 MHz)

Timebase

10 MHz In/Out (phase locks the internal frequency to other SR865s.)

Input impedance

1 MΩ or 50 Ω

Phase setting resolution

(360/232) degrees

Phase noise

 

     Int. reference

<0.0001° rms at 1 kHz (100 ms, 12 dB/oct)

     Ext. reference  (typ)

<0.001° rms at 1 kHz (100 ms, 12 dB/oct)

Phase drift (typ)

 

Sine Out to signal In (200 mVrms) 
<0.002°/°C below 20 kHz (DC coupled)
<0.02°/°C below 200 kHz
<0.2°/°C below 2 MHz

Harmonic detection

Detect at N × fref where N<99 and (N × fref ) < 2 MHz

Dual F reference

Detect at fdual =│fint - fext
All frequencies must be less than 2 MHz

Chopper reference

SR865 drives SR540 Chopper (via Aux Out 4) to lock the chopper to fint

Demodulator

DC stability

Digital output values have no offset drift

Time constants

1 µs to 30 ks

Low pass filters

Typical RC-type filters or advanced Gaussian/Phase-Linear filters

Filter slope

6, 12, 18 or 24 dB/oct rolloff

Synchronous filter

Available below 4 kHz

Harmonic rejection

-80 dB

Low latency output

Rear-panel BlazeX output with <2 µs delay (plus LPF rise/fall times).

Internal Oscillator

Frequency

mHz to 2 MHz, specified (operates to 2.5 MHz)

Frequency accuracy

25 ppm + 30 µHz (with internal timebase)

External timebase

10 MHz timebase input/output on rear panel

Frequency resolution

6 digits or 0.1 mHz (whichever is greater)

Sine Output

Outputs

Differential or single-ended

Output impedance

50 Ω source

Amplitude

nVrms to 2 Vrms (specified amplitude is differential into 50 Ω loads)
Output amplitude is halved when used single-ended
Output amplitude is doubled into a high impedance load

Amplitude resolution

3 digits or 1 nV, whichever is greater

DC offset

±5 V, differential or common mode

Offset resolution

3 digits or 0.1 mV, whichever is greater

Output limit

±6 V, sum of DC offset and peak amplitude

Sync

Logic level sync on rear panel (via BlazeX output)

Data

Data channels

4 data channels are displayed and graphed (green, blue, yellow, orange)

Data sources

Each data channel can be assigned any of these data sources:
X, Y, R, Θ, Aux In 1 to 4, Aux Out 1 to 2, X noise, Y noise, Sine Out Amplitude, 
Sine Out DC Level, Reference Phase, or Reference Frequency

Data history

All data sources are continually stored at all chart display time scales.
The complete stored history of any data source can be displayed at any time.

Offset

X, Y and R may be offset up to ±999 % of the output scale

Expand

X, Y and R may be expanded by ×10 or ×100

Ratio

X and Y may be ratioed by Aux In 3. R may be ratioed by Aux In 4

Capture buffer

Mpoints internal data storage. Store (X), (X and Y), (R and Θ) or (X, Y, R and Θ)
at sample rates up to 1.25 MHz. This is in addition to the data histories for the chart display.

Data streaming

Realtime streaming of data, either (X), (X and Y), (R and Θ), or (X, Y, R and Θ) at sample rate to 1.25 MHz over Ethernet interface

Scanning

One of the following parameters may be scanned:
fint, Sine Out Amplitude, Sine Out DC Level, Aux Out 1 or 2.

FFT

Source

Input ADC, demodulator output, or filter output

Record length

I1024 bins

Averaging

Exponential rms

Inputs and Outputs

CH1 output

Proportional to X or R, ±10 V full scale thru 50 Ω

CH2 output

Proportional to Y or Θ, ±10 V full scale thru 50 Ω

X and Y rear-panel outputs

Proportional to X and Y, ±10 V full scale thru 50 Ω

BlazeX

Low latency output of X, ±2.0 V full scale or logic level reference sync output, either thru 50 Ω

Aux outputs

4 BNC D/A outputs, ±10.5 V thru 50 Ω, 1 mV resolution

Aux inputs

4 BNC A/D inputs, ±10.5 V, 1 mV resolution, 1 MΩ input

Trigger input

TTL input triggers storage into the internal capture buffer

Monitor output

Analog output of the signal amplifier

HDMI

Video output to external monitor or TV, 640 × 480, 60 Hz

Timebase I/O

Vrms, 10 MHz clock to synchronize internal reference frequency to other units

General

Interfaces

IEEE488.2, RS-232, USB device and Ethernet.

USB flash

Front-panel slot for USB flash storage of screen shots and data

Preamp power

9-pin D connector to power SRS preamps

Power

60 W, 100/120/220/240 VAC, 50/60 Hz

Dimensions

17" × 5.25" × 17" (WHL)

Weight

22 lbs.

Warranty

One year parts and labor on defects in materials and workmanship